The new version 1.1 of gFM-Dental, the tool for the daily and monthly constancy testing of diagnostic monitors of digital X-ray devices for dentists, now supports the new DIN 6868-157 standard. The new standard was published on November 1, 2014 and includes acceptance and constancy testing for quality assurance measures on image display systems and their environment. This is intended to ensure that the entire image display chain is tested. For the implementation of the new directive, room classes have been introduced that take into account the illuminance of the ambient light and the area of activity in the requirements. The new DIN 6868-157 standard applies to all monitors put into operation from May 1, 2015. A transitional period of ten years applies to all monitors put into operation and successfully approved before this date. Find out in this article what will change in the daily constancy test for devices from May 1, 2015.
Daily constancy check
As before, a diagnostic monitor must also be checked daily for correct function under the new DIN 6868-157 standard. The daily constancy test is a visual check on the new TG18-UIQ test image, which involves checking a total of four criteria in the dental practice. The new test image is included with gFM-Dental and can be set up in the settings as the test image to be displayed.
The following criteria must be checked during the daily constancy test in the dental practice:
- Unadulterated visibility of the line pair grids without smearing, blurring or erasure and visibility of the two-pixel-wide line pair grids with low contrast in the center and in all four corners
- Visibility of the borders and lines of the main grid and centering of the grid in the display area
- Continuity of the appearance of the Gray gradient bar
- Visibility of the 95% and the 5% field within the larger fields
- Room classes 1-3 and mammography: The words "QUALITY CONTROL" with low contrast must be completely legible in the gray and white fields. Depending on the type of use and room class, different numbers of letters of the words must be legible (does not apply to RK5 and RK6 in dentistry).
Annual constancy test
The monthly constancy test is no longer required under the new DIN 6868-157 standard. In dentistry with room classes 5 and 6, the semi-annual constancy test only has to be carried out annually. In the annual constancy test, two visual tests are carried out: the homogeneity within an image reproduction device and a check of the color impression and uniformity. The luminance is determined for 18 gray value levels and it is checked whether the DICOM characteristic curve is displayed correctly. In addition, the minimum and maximum luminance is checked, from which the maximum luminance ratio is determined.
The following criteria are checked during the annual constancy test in dentistry (RK5 and RK6):
- Minimum luminance (metrological method)
- Maximum luminance (metrological method)
- Maximum luminance ratio (Calculation from minimum and maximum luminance)
- Illuminance (metrological method)
- Homogeneity of the image reproduction device (visual with test image)
- Color impression and uniformity (visual with test image)
Further test criteria are required for other room classes outside dentistry.
Daily constancy check with gFM-Dental
You can set up the required test image in gFM-Dental to perform the daily constancy test of a findings monitor that was put into operation and successfully approved after May 1, 2015.
The test image and all necessary test criteria are set up in gFM-Dental under the menu item [Settings]. Under the item [Screen resolution] the entry [DIN 6868-157 (as of May 1, 2015) off. Remove the test criteria for the daily constancy test from gFM-Dental by clicking on the red [Remove]-button to the left of the entry. Now add the data described in this article in the section "Daily constancy test" mentioned four test criteria in dentistry.
Once you have set up gFM-Dental as described, you can carry out the daily constancy test in accordance with the new DIN 6868-157 standard.
For existing customers, an update to gFM-Dental 1.1 is generally not necessary, as screens that were put into operation before May 1, 2015 do not have to be checked in accordance with the new DIN 6868-157 standard.
Test gFM-Dental without obligation
The current version gFM-Dental 1.1 is now available for download. Download and can be tested free of charge and without obligation for a period of 30 days without a registration key. A valid license key can be obtained from the goFileMaker online store can be purchased.
Did you know that you can customize and extend gFM-Dental with additional functions to meet your individual requirements? gFM-Dental is based on the FileMaker Pro 13 database system with the gFM-Business platform and can easily be extended with additional functions. The ability to control and read third-party hardware enables a wide range of applications such as employee time recording, document archiving with scanner support or the automated documentation of devices such as sterilizers, etc.
Apple Mac OS X in the dental practice
More and more dentists are opting to use Apple computers in the dental practice. The advantages are obvious: in addition to optical aspects, the Apple Macintosh is considered to be extremely reliable and low-maintenance. Due to its high value retention, Apple hardware is also considered very cost-effective in the business environment. The IT office Dirk Küpper from Düsseldorf has already completely converted more than 20 dental practices to Apple Mac OS X. If you are interested, Dirk Küpper can provide you with expert advice on the use of Apple computers in your dental practice.
[wpshopgermany product="22″]

Markus Schall has been developing individual databases, interfaces and business applications based on Claris FileMaker since 1994. He is a Claris partner, FMM Award winner 2011 and developer of the ERP software gFM-Business. He is also a book author and founder of the M. Schall Publishers.